Home  /  News

Introduction of electronic colposcopy manufacturers: What problems are prone to occur during the use of electronic colposcopy

Released on Mar. 11, 2021

Common cervical cancer screening examination items include colposcopy. Under colposcopy, select suspicious cancerous areas for cervical biopsy. Whether it is colposcopy or other equipment, if it is used for a long time, various problems will always occur. The following electronic Introduction of colposcopy manufacturers: What Problems Are Prone To Occur During The Use Of Electronic Colposcopy? Help doctors better understand and operate machines.


The following problems are prone to occur during the use of electronic colposcopy:

1. Various problems of the display:

1) The color monitor does not display

Mainly analyze the following reasons: 1. The monitor switch is not turned on; 2. The computer power is not turned on; 3. Whether the power socket is plugged in; 4. The monitor is broken;

2) After the host is turned on, the indicator light of the display is orange

Mainly analyze the following reasons: 1. The motherboard is broken; 2. The memory is broken; 3. Re-plug or wipe the memory with alcohol;

3) Fuzzy display

Mainly analyze the following reasons: 1. The graphics card is broken; 2. The graphics card is in poor contact; 3. Magnetic field interference; 4. The display signal line is poor in contact; 5. The true color setting of the display properties is incorrect;

4) Display ghosting

Mainly analyze the following reasons: 1. Poor focus; 2. Magnetic field interference;

5) Display color cast

Mainly analyze the following reasons: 1. The graphics card is broken; 2. The display is broken; 3. Magnetic field interference;


2. There is no image in the image display area

Mainly analyze the following reasons: 1. Whether the power of the camera is properly connected; 2. Whether the video cable of the camera is well connected to the computer; 3. Whether the camera is turned on in the software;


3. the problems involved in the colposcopy camera

1. The cold light source is not bright:

Mainly analyze the following reasons: 1. First check whether the power supply is normal, check the fuse, and check the porcelain socket; 2. Replace the bulb or change the cold source.

1. The problem of zooming in or out

The colposcope mirror cannot be enlarged or reduced:

Mainly analyze the following reasons: 1. The line is not in contact with the host; 2. The COM port is not set up; 3. Press the zoom button behind the mirror at the same time;

The zoom in and zoom out function of the lens cannot be controlled:

The main reason for the analysis: the circuit board in the control panel of the colposcopy is broken;

The software's zoom-in and zoom-out functions cannot be controlled:

The main reason for the analysis: the COM port setting is wrong;


4.about printing reports

Because the report needs to be printed out, it is mainly convenient for the doctor to check the disease and save the file for later checking and printing. The following problems often occur:

1.Print color cast

Mainly analyze the following reasons: 1. Ink shortage; 2. Clean the print head; 3. Replace the empty cartridge; 4. If it is a new ink cartridge, the ink cartridge has quality problems;

2. Every time the printer is turned on, a piece of paper is empty, and it will be normal in the future

Mainly analyze the following reasons: 1. The printer is caused by a sudden power failure; 2. Adjust the printer properties;

3. The printer stops when printing halfway, indicating that the signal line is wrong

Mainly analyze the following reasons: 1. Reinstall the printer driver; 2. Change a signal line;

4. Paper jam or no printing during printing, prompting out of paper

Mainly analyze the following reasons: 1. The resistance of the ink cartridge guide rail is too large, wipe the slide bar with sewing machine oil; 2. There is torn paper or other debris in the printer that is stuck;

5. The printer is normal, but does not print

Mainly analyze the following reasons: 1. The printing cable is broken; 2. The printer interface is broken; 3. The motherboard interface is broken; 4. The driver is incorrect

6. The printer prints garbled characters

Mainly analyze the following reasons: 1. Reinstall the driver; 2. Clean the print head; 3. Turn off the printer and restart; 4. The printer is broken; 5. Replace the printing cable

7. Pay attention when placing the paper, put it in the paper feed slot vertically, and clamp it with the clips on both sides, otherwise the report will be deflected and skewed.

8. If there is a paper jam in the printer, first turn off the workstation, carefully pull open the back cover, and take out the jammed paper from the inside.

9. After the toner of the laser printer is used up, the entire toner cartridge must be replaced. The toner cannot be replaced, otherwise the printer will leak toner and damage the printer.